Analog Programming (APG) and Isolated Analog Programming (ISOL)

4-28

975-0200-01-01

Deactivating ISOL Current Mode

To deactivate the ISOL current mode:

1. Turn the mode control knob to the CAP position or press the Rotary

knob/Enter button if the control knob is already at the CAP position.

2. CAPr is displayed on the output voltage display.

3. Turn the Rotary knob/Enter button to select nonE.

4. Press the Rotary knob/Enter button to commit the change.

The display will return to the normal display.

Query for ISOL Current Input Level

To query for ISOL current input level from non-isolated input:

1. Turn the mode control knob to the CAP position or press the Rotary

knob/Enter button if the control knob is already at the CAP position.
CAPr is displayed on the output voltage display.

2. If the mode is currently nonE then select any other ISOL mode. If

this step is done, be sure to restore the ISOL setting to nonE when
the operation is complete. Otherwise, do not rotate the Rotary knob/
Enter button to keep the current ISOL mode.

3. Press the Rotary knob/Enter button to select the ISOL mode.

The output voltage display should read LE C and the output current
display will display ISOL level.

4. Press the Rotary knob/Enter button to return to normal mode.

5. If the ISOL mode was modified in step 2, execute the procedure to

deactivate the ISOL voltage mode.

Quick Tip

Remote operation
uses these SCPI
commands. See
page 5–1
.

The SCPI command (s) for these instructions are:

[:]SYSTem[<channel>]:REMote:CURRent:SOURce:ISOlated

[:]SYSTem:COMMunicate:APRogram:CURRent[:ISOlated]:LEVel

[:]MEASure:APRogram:CURRent:ISOlated[:SCALar][:DC]?