Multichannel address setting, Remote interface addressing – Xantrex Technology XTR 850 Watt User Manual

Page 143

Advertising
background image

Terminal Configuration

975-0200-01-01

5-21

5

Multichannel

Address Setting

The power supply multichannel address can be set to any address between
1 to 30. All units that are connected together via the RS-232 or RS-485
connector must have a unique multichannel address.

To set the address:

1. Select the appropriate communication port as described “Selecting

the Appropriate Communication Port” on page 5–20.

If more than one unit has the same multichannel address when
connected through the RS-485 connector, then collisions are possible,
resulting in garbled responses to queries.

2. Press the Rotary knob/Enter button.

ADDR is displayed on the output voltage display.

3. Turn the Rotary knob/Enter button to select the desired multichannel

address

between 1 to 30.

4. Press the Rotary knob/Enter button to commit the new address.

Remote Interface Addressing

All commands must be issued with a multichannel address or the device
must be selected using the:

*adr or :SYST[<channel>]:COMM[:MCH]:ADDR commands.

Once a device is selected all commands sent without a multichannel
address will be handled by the selected device. The use of multichannel
addresses supersedes the selected device as the destination for a message.
For more details on how multichannel addresses affect responses, see
Table 5-8.

To change the selected device simply issue the *adr or
:SYST[<channel>]:COMM[:MCH]:ADDR

command with a

different multichannel address. The previously selected device will be
deselected and the new device will be selected. From that point on, all
messages without a multichannel address will be destined for the newly
selected device.

The SCPI Commands for these instructions are:

[:]SYSTem:COMMunicate[:MCHannel]:ADDRess <address>

or

*ADR <address>