Calibration procedure for isolated modes, Isolated voltage monitoring calibration, Isolated current monitoring calibration – Xantrex Technology XTR 850 Watt User Manual

Page 201: Calibration procedure for isolated modes –13

Advertising
background image

Calibration and Troubleshooting

975-0200-01-01

6-13

6

Calibration Procedure for Isolated Modes

The main output and the non-isolated mode must be calibrated first.

Isolated Voltage Monitoring Calibration

To calibrate the isolated voltage monitoring:

1. Disconnect the load and connect the voltmeter to the output.

2. Connect the voltmeter to the isolated voltage monitoring output lines

(J3.5 - J3.6).

3. Set the Voltage APG level to 4 V. The SCPI command is:

SYST:COMM:APR:LEV 4

4. Set the input source to local for voltage analog programming. The

SCPI command is:

SYST:REM:SOUR:ISOL LOC

5. Set the main output voltage to 100% by sending the command:

SOUR:VOLT <ModelVolt>

or use the Rotary knob/Enter button.

6. Turn the main output to On by pressing the OUTPUT ON/OFF button

on the front panel or send the command:

:OUTP ON

The OUTPUT ON/OFF button will illuminate.

7. Enter the interactive calibration mode by sending the SCPI command:

CAL:MON:ISOL

8. Press the terminal keys – or + to reduce or increase the monitoring

value. Adjust the monitoring value displayed on the voltmeter that
was setup in Step 1 to read as close as possible to 4.000 V.

9. Press any key other than the + or – keys to exit from interactive

calibration mode.

Isolated Current Monitoring Calibration

To calibrate the isolated current monitoring:

1. Short the main output with a shunt. Connect a multimeter to measure

the voltage across the shunt.

2. Connect the voltmeter to the Isolated monitoring output lines (J3.10

and J3.6).