Calibration and Troubleshooting

6-10

975-0200-01-01

5. Set the input source as Voltage Non-Isolated for Voltage

Programming. The SCPI command is:

SYST:REM:SOUR AVOL

6. Turn the main output to On by pressing the OUTPUT ON/OFF button

on the front panel or send the command:

:OUTP ON

The OUTPUT ON/OFF button will illuminate.

7. Enter the interactive calibration mode by sending the SCPI command:

CAL:OUTP:ANAL

8. Press the terminal keys – or + to reduce or increase the main output

voltage value. Adjust the main output voltage to 100%, measured on
the external voltmeter on the output.

9. Press any key other than the + or – keys to exit from interactive

calibration mode.

Non-isolated Resistive Programming of Voltage Calibration

To calibrate the non-isolated resistive programming of voltage:

1. Disconnect the load and connect the voltmeter to the output.

2. Connect the 4.000k

Ω resistor to the non-isolated connector, resistive

programming of voltage, and voltage programming (J1.9-J1-11, and
J1.7).

3. Set the resistive APG level to 4k

Ω. The SCPI command is:

SYST:COMM:APR:LEV 4

4. Set the input source as Resistive Non-Isolated for Voltage

Programming. The SCPI command is:

SYST:REM:SOUR ARES

5. Turn the main output On by pressing the OUTPUT ON/OFF button

on the front panel or send the command:

:OUTP ON

The OUTPUT ON/OFF button will illuminate.

6. Enter the interactive calibration mode by sending the SCPI command:

CAL:OUTP:ANAL:RES

7. Press the terminal keys – or + to reduce or increase main output

voltage value. Adjust the main output voltage to 100%.