Xantrex Technology XTR 850 Watt User Manual

Page 202

Advertising
background image

Calibration and Troubleshooting

6-14

975-0200-01-01

3. Set the current APG level to 4 V. The SCPI command is:

SYST:COMM:APR:LEV:CURR 4

4. Set the input source to local for voltage analog programming. the

SCPI command is:

SYST:REM:SOUR:CURR:ISOL LOC

5. Set the main output voltage to 100% by sending the command:

SOUR:VOLT <ModelVolt>

or use the Rotary knob/Enter button.

6. Turn the main output to On by pressing the OUTPUT ON/OFF button

on the front panel or send the command:

:OUTP ON

The OUTPUT ON/OFF button will illuminate.

7. Enter the interactive calibration mode by sending the SCPI command:

CAL:MON:CURR:ISOL

8. Press the terminal keys – or + to reduce or increase the monitoring

value. Adjust the monitoring value displayed on the voltmeter that
was set up in Step 1 to read as close as possible to 4.000 V.

9. Press any key other than the + or – keys to exit from interactive

calibration mode.

Isolated Voltage Programming of Voltage Calibration

To calibrate the isolated voltage programming of voltage:

1. Disconnect the load and connect the voltmeter to the output.

2. Connect the input voltage source to the Isolated connector (J3.3 and

J3.6).

3. Set the Voltage APG level to 4 V. The SCPI command is:

SYST:COMM:APR:LEV 4

4. Set the input source voltage to 4.000 V.

5. Set the input source as Voltage Isolated for Voltage. The SCPI

command is:

SYST:REM:SOUR:ISOL AVOL

6. Turn the main output to On by pressing the OUTPUT ON/OFF button

on the front panel or send the command:

:OUTP ON