Calibration and Troubleshooting

6-16

975-0200-01-01

Isolated Voltage Programming of Current Calibration

To calibrate the isolated voltage programming of current:

1. Short the main output with a shunt. Connect a multimeter to measure

the voltage across the shunt.

2. Connect the input voltage source to the Isolated connector (J3.4 -

J3.6).

3. Set the Current APG level to 4 V. The SCPI command is:

"SYST:COMM:APR:LEV:CURR 4"

4. Set the input source voltage to 4.000 V.

5. Set the input source as Voltage Isolated for Current. The SCPI

command is:

SYST:REM:SOUR:CURR:ISOL AVOL

6. Turn the main output to On by pressing the OUTPUT ON/OFF button

on the front panel or send the command:

:OUTP ON

The OUTPUT ON/OFF button will illuminate.

7. Enter the interactive calibration mode by sending the SCPI command:

CAL:OUTP:ANAL:CURR:ISOL

8. Press the terminal keys – or + to reduce or increase the main output

current value. Adjust the main output current to 100%.

9. Press any key other than the + or – keys to exit from the interactive

calibration mode.

Isolated Resistive Programming of Current Calibration

To calibrate the isolated resistive programming of current:

1. Short the main output with a shunt. Connect a multimeter to measure

the voltage across the shunt.

2. Connect the 4.000 k

Ω resistor to the isolated connector. One terminal

to resistive programming of current (J3.7), and voltage programming
of current (J3.3), and the other to common (J3.2).

3. Set the Resistive APG level to 4 k

Ω. The SCPI command is:

SYST:COMM:APR:LEV:CURR 4

4. Set the input source as Resistive Isolated for Current Programming.

The SCPI command is: