Calibration and Troubleshooting

6-2

975-0200-01-01

Introduction

The calibration of the power supply is software dependent, and there are
no potentiometers to adjust.

Calibration is performed via SCPI commands.

The following items need to be calibrated:

Programmed voltage

Voltage readback

Programmed current

Current readback

Non-isolated voltage monitoring

Isolated voltage monitoring

Non-isolated current monitoring

Isolated current monitoring

Isolated source for analog programming

Non-isolated source for analog programming

Over voltage protection system

Main Voltage and Current Calibration Principle

Understanding the Problem

Figure 6-1 illustrates two sources of analog programming error: gain error
and offset error. Gain error is the departure from the ideal slope of the
measured versus programmed line. Offset error is the magnitude of the
measured value when the programmed value is zero.

For best results, the next recommended action is gain calibration for 90%
of maximum value and offset calibration for 10% of maximum value.
Perform gain calibration again.

CAUTION

Calibration procedures should only be performed by qualified user.