Calibration and Troubleshooting

6-8

975-0200-01-01

Non-isolated Analog Programming Calibration

Prior to this, the main output must be calibrated first.

Non-isolated Voltage Monitoring Calibration

To calibrate the non-isolated voltage monitoring:

1. Disconnect the load and connect the voltmeter to the output.

2. Connect the voltmeter to the non-isolated voltage monitoring output

lines (J1.13 - J1.7).

3. Set the Voltage APG level to 4 V. Send the SCPI command:

SYST:COMM:APR:LEV 4

4. Set the input source as Local. Send the SCPI command:

SYST:REM:SOUR LOC

5. Set the main output voltage to 100% by sending the command:

SOUR:VOLT <ModelVolt>

or use the Rotary knob/Enter button.

6. Turn the main output On by pressing the OUTPUT ON/OFF button

on the front panel or send the command:

:OUTP ON

The OUPTUT ON/OFF button will illuminate.

7. Enter interactive calibration mode by sending the SCPI command:

CAL:MON

8. Press on the terminal keys – or + to reduce or increase monitoring

value. Adjust the monitoring value displayed on the voltmeter that
was set up in Step 1 to read as close as possible to 4.000 V.

9. Press any key other than the + or – keys to exit from interactive

calibration mode.

Important:

In calibration commands, when + or – keys are expected, any

other key will exit from the calibration mode.