Calibration and Troubleshooting

975-0200-01-01

6-11

6

8. Press any key other than the + or – keys to exit from interactive

calibration mode.

Non-isolated Voltage Programming of Current Calibration

To calibrate the non-isolated voltage programming of current:

1. Short the main output with a shunt. Connect a multimeter to measure

the voltage across the shunt.

2. Connect the input voltage source to the non-isolated connector,

current programming (J1.10 - J1.7).

3. Set the Voltage APG level to 4 V. The SCPI command is:

SYST:COMM:APR:LEV:CURR 4

4. Set the input source voltage to 4.000 V.

5. Set the input source as Voltage Non-Isolated for Current. The SCPI

command is:

SYST:REM:SOUR:CURR AVOL

6. Turn the main output On by pressing the OUTPUT ON/OFF button

on the front panel or send the command:

:OUTP ON

The OUTPUT ON/OFF button will illuminate.

7. Enter the interactive calibration mode by sending the SCPI command:

CAL:OUTP:ANAL:CURR

8. Press the terminal keys – or + to reduce or increase the main output

current value. Adjust the main output current to 100%.

9. Press any key other than the + or – keys to exit from interactive

calibration mode.

Non-isolated Resistive Programming of Current Calibration

To calibrate the non-isolated resistive programming of current:

1. Short the main output with a shunt. Connect a multimeter to measure

the voltage across the shunt.

2. Connect the 4.000 k

Ω resistor to the non-isolated connector (APG -

J1), one terminal to resistive programming of current (J1.12) and
voltage programming of current (J1.10), and the other terminal to the
common (J1.7).

3. Set the resistive APG level to 4 k

Ω. The SCPI command is: