Non-isolated current monitoring calibration – Xantrex Technology XTR 850 Watt User Manual

Page 197

Advertising
background image

Calibration and Troubleshooting

975-0200-01-01

6-9

6

Non-isolated Current Monitoring Calibration

To calibrate the non-isolated current monitoring:

1. Short the main output with a shunt. Connect a multimeter to measure

the voltage across the shunt.

2. Connect the voltmeter to the non-isolated current monitoring output

lines (J1.12 - J1.7).

3. Set the Current APG level to 4 V. Send the SCPI command:

SYST:COMM:APR:LEV:CURR 4

4. Set the input source as Local.

5. The SCPI command is:

SYST:REM:SOUR:CURR LOC

6. Set the main output voltage to 100% by sending the command

SOUR:VOLT <ModelVolt> or use the Rotary knob/Enter button.

7. Turn the main output On by pressing the OUTPUT ON/OFF button

on the front panel or send the command:

:OUTP ON

The OUTPUT ON/OFF button will illuminate.

8. Enter interactive calibration mode by sending SCPI command:

"CAL:MON:CURR"

9. Press on the terminal keys – or + to reduce or increase the monitoring

value. Adjust the monitoring value displayed on the voltmeter that
was set up in Step 1 to read as close as possible to 4.000 V.

10. Press any key other than the + or – keys to exit from interactive

calibration mode.

Non-isolated Voltage Programming of Voltage Calibration

To calibrate the non-isolated voltage programming of voltage:

1. Disconnect the load and connect the voltmeter to the output.

2. Connect the input voltage source to the non-isolated connector,

voltage programming (J1.9 - J1.7).

3. Set the Voltage APG level to 4 V. Send the SCPI command:

SYST:COMM:APR:LEV 4

4. Set input source voltage to 4.000 V.