Caution, Tiis, Connecting cables to the protection system – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 135

Advertising
background image

<2. Installation, Piping, and Wiring>

2-18

IM 11B08A01-01E

TIIS

For TIIS explosionproof wiring connections, use cable packing adapters or sealing fi ttings (for
Ethernet cable).

A

Connection port of the protection system

Connection port of the electronics section *1

Connector *4

Cable
packing adapter

Nipple

Gasket

Nut

Detail of A

Sealing fitting
(Accessory kit)
*2

Electronic section
(Pressurized enclosure)

Protection system

(Flameproof enclosure)

Cable packing adapter
(Accessory kit) *3

*1: Any of the six connection ports of the electronics

section can be equally used.

*2: To mount the sealing fitting, remove the cable packing

adapter from one of the six connection ports of the
electronics section.

*3: To use the protection system, remove the plug from its

connection port and mount the cable packing adapter.

*4: A connector (3/4NPT) can be mounted on any of the

connection ports.

Figure 2.12

Wiring in TIIS specifi cation

Connecting cables to the protection system

Remove the attached plug and mount the cable packing adapter (G3/4) in the accessory kit
on the connection port. Use the connector in the accessory kit for the cable packing adapter
(3/4NPT).

CAUTION

The minimum packing (for ø8.0 to ø9.0) is attached to the cable packing adapters at shipment.
Change it to an appropriate packing for the cable O.D. (See Table 2.6.)

2nd Edition : May 11, 2012-00