Peak copy 2 – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 323

Advertising
background image

<5. EtherLCD>

5-72

IM 11B08A01-01E

4. Copy rules for synchronized peaks.

Destination

Source

Independent peak

Parent peak

Child peak

Independent peak Copy as is (change copy

item)

Maintain destination peak

in independent peak and

synchronize (comply with

synchronization rules to

change synchronization

items).

Assess using peak

synchronizing rules. When

synchronization is not possible,

maintain independent peak

and synchronize. When

synchronization is possible,

set synchronization as

synchronization target of copy

destination = synchronization

target of copy source.

Parent peak

Cannot copy

Cannot copy

Cannot copy

Child peak

Convert to independent

peak and copy (change

copy item).

Convert to independent

peak and synchronize

(comply with

synchronization rules to

change synchronization

items).

Assess using peak

synchronizing rules. When

synchronization is not possible,

maintain independent peak

and synchronize. When

synchronization is possible,

set synchronization as

synchronization target of copy

destination = synchronization

target of copy source.

5. Copy items: Copy stream number in peak process-specifi c setup part (T19) and parts other

than relative peak number

Peak copy 2

Press F3 (Copy 2).

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Peak Copy 2

(1/2)

Stream # 1 Use num # 1 Assign #

10

Peak# Name Usage Process (Detail)
1 P1

On Base level

2 Off Indirect
3 Off Indirect
4 Off Indirect
5 Off Indirect
6 Off Indirect
7 Off Indirect
8 Off Indirect

Menu PeakSet

Copy

F1

F2

F3

F4

F5

F6

Figure 5.80

Example of peak copy 2 screen

F1 (Menu):

Displays the Table Menu screen.

F2 (PeakSet): Displays the Peak Setup-Specifi c screen
F6 (Copy):

Executes copy after entering Source peak and Destination peak.

• Stream number for display and setting becomes the stream number in the peak setup-

specifi c screen when relevant screen is requested.

• Copy destination peak must be an independent peak.
• See Peak synchronization and peak copy items regarding data for copy.
• When Process (Type) is set to No processing, Process (Detail) displays No processing.

2nd Edition : May 11, 2012-00