App.e-3, 3) signal evaluation, Process mode processing – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 476: App.e

Advertising
background image

<Appendix E Computation Scheme>

App.E-3

IM 11B08A01-01E

Waveform of detector signals

Waveform after filtering
f=0.5

Waveform after sampling
(t

smp

=40ms)

Waveform after sampling
(t

smp

=80ms)

F020202.ai

(3) Signal Evaluation

The signal evaluation step calculates the standard deviation of sampled 20 signals according to

the following formulas:

y

avrg

=

y

i

1

20

t

i=t–19

s

d

=

(y

i

– y

avrg

)

2

1

20

t

i=t–19

Where:

y

i

: sampled signals (20 signals)

s

d

: standard deviation of the sampled data

Process Mode Processing

There are four kinds of process mode:

1. Signal level processing
2. Peak processing
3. Deviation component computation processing
4. Additional computation processing

Up to 255 settings can be computed and they are separately divided for each stream.

2nd Edition : May 11, 2012-00

App.E

Advertising