4 detector setup – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 299

Advertising
background image

<5. EtherLCD>

5-48

IM 11B08A01-01E

5.4.4 Detector

Setup

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Detector Setup

(1/2)

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Detector Setup

(2/2)

Oven #

1 Det # 1 ( TCD )

Oven #

1 Det # 1 ( TCD )

>

- Filtering const Off

>

- Auto gain setting None

- Sample rate 20 ms

- Full scale value 0

- Square root calc req * * * * *

- Auto zero Off

- Flame detect level * * * * *

- Auto zero value 0.0000 mV

- Sense set Equal

- Auto zero time 0.0 s

- Input resistor * * * * * * * * * *

Menu

Oven#

Det#

Menu

Oven#

Det#

F1

F2

F3

F4

F5

F6

F1

F2

F3

F4

F5

F6

Figure 5.62

Example of the detector setup screen

F1 (Menu):

Displays the Table Menu screen.

F5 (Oven #): Sets an oven number.
F6 (Det #):

Sets a detector number.

TIP

• The oven numbers set to Not provided on the Temperature Control Set cannot be selected.
• If the Detector type has been set to None on the Detector Set screen, all the options are shown as *

(asterisk).

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

Oven #

Oven number

1

3

Det #

Detector number

1

2

Detector type

None

TCD

FID

FID-MC

FPD

Filtering const

0.001

1.000

Sample rate

○ *

20 ms

40 ms

80 ms

160 ms

Square root

calc req

The value is shown except

when the FPD is installed.

No

Yes

* The square root of the area

value

Flame detect

level

–20

20

mV

The item is displayed only when

the FID, FPD or FID-MC is

installed.

Sense set

No amplifi cation

Tenfold

Input resistor

1 Mohm

6 Mohm

20 Mohm

60 Mohm

200 Mohm

600 Mohm

1 Gohm

2 Gohm

*:

The settings are valid after rebooting or turning off/on its power.