Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 217

Advertising
background image

<4. GC-HMI (touch panel)>

4-30

IM 11B08A01-01E

Cal (Val):

Executes calibration or validation of the specifi ed number. After the
selection, set the calibration (validation) number (1 to 3 for calibration, or
1 to 3 for validation). Ensure that the calibration (validation) number is the
stream number.

Start cal (val):

Starts the calibration (validation).

Finish cal (val):

Finishes the calibration or validation.

Cancel spec.:

Cancels the parameter settings for changing the measurement status.

Cal (val) method:

Set the desired mode of calibration or validation: manual, semi-auto, or
auto.

Figure 4.36

Auto:

Calibration or validation is executed automatically according to the set
values of start time and time interval.

Semi-auto:

When a calibration or validation number is specifi ed, the stream valve
on the calibration or validation stream automatically opens, followed by
calibration or validation.

Manual:

When a calibration or validation number is specifi ed, all stream valves
close, allowing a standard sample to fl ow. Then, calibration or validation
can be performed manually.

Start/stop auto cal: Starts/stops automatic calibration.

2nd Edition : May 11, 2012-00