<Appendix D Action of External Input and Output Signals>

App.D-12

IM 11B08A01-01E

Table 10 Actions of D/O1

In the status of stream

sequence 1

Stop to Pause

Always On for the specifi ed stream sequence

regardless of the mode

Stop to Run
Run to Pause
Run to Stop

Run mode of stream sequence 1 to Run mode of

stream sequence 2

Turned Off for stream sequences other than the

specifi ed stream sequence regardless of the mode

Pause mode of stream sequence 1 to Stop mode

of stream sequence 2
Stop mode of stream sequence 1 to Stop mode of

stream sequence 2
Run mode of stream sequence 2 to Run mode of

stream sequence 1

Turned On for the specifi ed stream sequence

regardless of the mode

Pause mode of stream sequence 2 to Stop mode

of stream sequence 1
Stop mode of stream sequence 2 to Stop mode of

stream sequence 1
Run mode of stream sequence 1 to Run mode of

stream specifi cation 1

Turned Off for stream sequences other than the

specifi ed stream sequence regardless of the mode

Pause mode of stream sequence 1 to Stop mode

of stream specifi cation 1
Stop mode of stream sequence 1 to Stop mode of

stream specifi cation 1
Run mode of stream sequence 1 to Calibration/

Validation

Turned Off when analysis starts for the calibration/

validation stream

Pause mode of stream sequence 1 to Calibration/

Validation

Turned Off when the warming up time starts for the

calibration/validation stream

Stop mode of stream sequence 1 to Calibration/

Validation

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

D/O Setup

(1/2)

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

D/O Setup

(2/2)

Slot #

1 DO # 1

Slot #

1 DO # 1

>

- Process assign Stream sequence

>

- Stream sequence 7 Off

- GCM number 1

- Stream sequence 8 Off

- Stream sequence 1 On
- Stream sequence 2 Off
- Stream sequence 3 Off
- Stream sequence 4 Off
- Stream sequence 5 Off
- Stream sequence 6 Off

Menu

Status

Slot#

DO#

Menu

Status

Slot#

DO#

F1

F2

F3

F4

F5

F6

F1

F2

F3

F4

F5

F6

Figure 15 Process Specifi cation: Stream Sequence (D/O Setting Screen on GC-HMI EtherLCD)

Stream

The Stream function of D/O is a function to send out information that a specifi ed stream is

in the run or pause mode. When a stream set on the D/O setting screen is in operation, the

contact output is turned On regardless of the measurement status (stream sequence or stream

specifi cation).

Actions in the case where the stream number 1 is specifi ed for the contact number 1 are

described below. For example, Table 11 describes the action of D/O1 under the condition of

stream sequence and calibration setting shown in Table 9.

2nd Edition : May 11, 2012-00