D/o action setting screen – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 311

Advertising
background image

<5. EtherLCD>

5-60

IM 11B08A01-01E

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

GCM #

GCM number

1

6

SYS #

SYS number

1

6

Method #

Method number

1

6

ATM valve

ATM valve number

0

3 − 2

0: None

1 to 3 Oven

1 to 2 ATM valve

1st to 3rd ON

time

−Warming

up time

Analysis

cycle − 2

seconds ATM valve time setup

Some entry data are invalid.

1st to 3rd

OFF time

seconds

D/O Action Setting Screen

Press F5 (DO).

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

D/O Action Setting

GCM #1 / SYS #1 / Method #1

> - 1st DO ON time * * * * * * * s
- 1st DO OFF time * * * * * * * s
- 2nd DO ON time

* * * * * * * s

- 2nd DO OFF time * * * * * * * s
- 3rd DO ON time

* * * * * * * s

- 2nd DO OFF time * * * * * * * s

Seq

Gate

Valve

AtmV

F1

F2

F3

F4

F5

F6

Figure 5.74

Example of the D/O action setting screen

F1(Seq):

Displays SYS Sequential Display screen.

F2(Gate):

Displays Gate of the action set screen.

F3(Valve):

Displays ON/OFF time screen of a valve (1st, 2nd & 3rd of the maximum No. 1
to No. 18).

F4(Atm):

Displays ON/OFF time screen of ATM valve (1st, 2nd & 3rd).

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

GCM #

GCM number

1

6

SYS #

SYS number

1

6

Method #

Method number

1

6

1st to 3rd ON

Time

0

Analysis

cycle − 2

seconds D/O time setup

Some entry data are invalid.

* When ON time is invalid, the

minimum OFF time is 0.

1st to 3rd

OFF Time

ON time

Analysis

cycle − 2

seconds

2nd Edition : May 11, 2012-00