<6. Maintenance>

6-58

IM 11B08A01-01E

F0250.ai

In case of K9193NT, NY

In case of K9193NU, NX

(2)

(2)

(1)

Air

Air

SV11

EXH

IN

IN

IN

OUT

SV11

EXH

IN

OUT

OUT2

OUT2

OUT1

OUT1

PV12

PV11

PV11

Figure 6.50

(6) Diaphragm of Pneumatic Valve for Atmospheric Pressure Balancing or

Stream Switching

(1) Remove the pipe for air.
(2) Remove the four screws on the upper section of the stream switching valve (see Figure

6.51) or the atmospheric-pressure balancing valve (see Figure 6.52) to remove the upper
section of the valve itself.

(3) Replace the diaphragm with a new one and reassemble the valve. Install the new

diaphragm in the same orientation. Exercise care to avoid losing the adjusting pin and
spring at the lower section of the valve.

F0251.ai

1

10

No. Part name

Part number Quantity

air in

4

2

3

11

7

5

6

9

8

K9192WB

K9192W

A

1 BASE

K9192WD

1 1

2 BODY

K9192WE

1 1

3 CAP

K9192WF

1 1

4 PISTON

K9192WG 1 1

5 O-RING

K9142QU

1 1

6 O-RING

L9817FK

1 1

7 SPRING

K9192WH

1 1

8 SPRING

K9035AW

1 1

9 PIN

K9192WK

1 1

10 DIAPHRAGM K9192WL 1

K9192WV

1

11 SCREW

Y9335JU

4 4

Figure 6.51

Stream Switching Valve

2nd Edition : May 11, 2012-00