3 status and operation mode – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 63

Advertising
background image

<1. Overview>

1-36

IM 11B08A01-01E

1.11.3

Status and Operation Mode

Process

Run mode

Pause mode

Stop mode

Manual

Run mode

Stop mode

Run command

Stop command

Pause command

Change the status

Run

command

Pause

command

Stop

command

Stop

command

Run command

The GC8000 has the following statuses. The change is possible only in the Stop mode.

Process

Normal measurement, calibration and validation

Manual

Manual operation

The GC8000 has the following operation modes and commands.

Run

Mode in which measurement is running.
Starts the measurement in the Process mode.
This operation can be made by operators with user level B or higher.
Starts to display chromatograms in the Manual mode. (It does not detect
peaks, calculate concentration, or save chromatograms.) This operation
can be made by operators with user level C or higher.

Pause

Mode in which measurement pauses.
Operates until the pause time specifi ed in the GCM method in the
Process mode. This operation can be made by operators with user level
B or higher.

Stop

Mode in which measurement stops.
Operates in the Process mode for the main cycle specifi ed in the GCM
method. This operation can be made by operators with user level B or
higher.
Stops the Run mode immediately in the Manual status. This operation
can be made by operators with user level C or higher.

Command

cancellation

Cancels the operation mode command that is in the operation queue in

the Process mode. This operation can be made by operators with user

level B or higher.

Forced stop

Stops the current operation mode forcibly in the Process mode. This

operation can be made by operators with user level C+ or higher.

Figure 1.16 shows the change of status and operation mode.

Process

Manual

Run mode

Pause mode

Stop mode

Pause command

(Revocable)

Run

command

Pause

command

Stop command

(Revocable)

Stop

command

Run command

Forced stop command

Forced stop command

Change the status

Run mode

Stop mode

Run command

Stop command

Figure 1.16

Diagram of Changing the Status and Operation Modes

The statuses and operation modes can be changed in the Analyzer Operation screen of the GC-
HMI (touch panel) equipped on the GC8000 analyzer. The operation modes can be changed in
the Analyzer Overview screen.

2nd Edition : May 11, 2012-00

Advertising