App.e-6, Peak detection, B) auto-tracking peak processing – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 479

Advertising
background image

<Appendix E Computation Scheme>

App.E-6

IM 11B08A01-01E

Peak area/peak height

F020502_2.ai

Select one of the following

Concentration computation processing

External standard

(approximation

from cubic

equation)

External standard

(approximation

from linear

equation)

Total

compensation

operation

Variation

coefficient

External standard

(calibration

computation)

Indirect method

Corrected area

normalization

Select wheter to execute

Concentration value

Variation coefficient

Peak Detection

The details on square processing, retention time detection processing, and tangent processing

are the same as in peak processing in Manual mode.

(a) Base Line Correction Processing

Base Line Correction

In base line correction, stored chromatogram data are subtracted from the measured detector

signal data, and processing following the peak detection are carried out on the subtracted data.

When storage is specifi ed, base line correction is not done.

F02040101.ai

Measured chromatogram

Stored chromatogram

Corrected chromatogram

You can select whether or not correction is carried out for each stream.

(b) Auto-tracking Peak Processing

As the retention time changes, gates for each peak are displaced. To prevent this displacement,

auto-tracking peak processing automatically changes the gate time for each peak according to

the change in the retention time.

2nd Edition : May 11, 2012-00