Run

W1

W1

S1

S1

[2F2]

Time

Stream specification 1

Forced stop command

Stop

● In the case where a forced stop command is received while the operation mode is

scheduled to be changed to the stop mode or pause mode in the stream specifi cation status
[2F3], [2F4]

The operation mode is immediately changed to the stop mode in the stream specifi cation

status.

In the stream specifi cation status (for N times of measurement), the number of times of

measurement is counted again from the beginning.

Run

(Waiting for stop or pause)

W1

W1

S1

S1

[2F3] [2F4]

Time

Stream specification 1

Forced stop command

Stop

● In the case where a forced stop command is received during the pause mode in the stream

specifi cation status [2F6]

The operation mode is immediately changed to the stop mode in the stream specifi cation

status.

In the stream specifi cation status (for N times of measurement), the number of times of

measurement is counted again from the beginning.

Pause

[2F6]

Time

Stream specification 1

Forced stop command

Stop

● In the case where a pause command is received during the stop mode in the stream

specifi cation status [2H1]

The operation mode is immediately changed to the pause mode in the stream specifi cation

status.

Stop

[2H1]

Time

Stream specification 1

Pause command

Pause