Sys sequential display screen – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 307

Advertising
background image

<5. EtherLCD>

5-56

IM 11B08A01-01E

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

GCM #

GCM number

1

6

SYS #

SYS number

1

6

Method #

Method number

1

6

Gas # − #

Carrier gas number

1

2

Provides two each to Ovens 1,

2 & 3.

Initial press

setpoint

0

0.0

490

71.0

kPa

psi

Pressure unit is set in the

operating parameters screen.

(see SI unit limitation)

Some data are invalid.

#1 Press

ramp start

time

0

Analysis

cycle

s

Some data are invalid.

EPC program

time

0

Analysis

cycle

Displays total time of oven

program.

Initial time setpoint + Σi (Press

hold setpoint i − Press hold

setpoint i − 1)/Press ramp rate +

Temp hold time)

* When the temp ramp rate is

zero, only temp hold time is

added.

* When data is all invalid,

displays program time as 0s.

1st to 3rd

press ramp

rate

0.0

490

kPa/

min

psi/min

Pressure unit is set in the

operating parameters screen.

(see SI unit limitation)

Some data are invalid.

1st to 3rd

press hold

setpoint

0.0 *1

0.0 *1

490

71.0

kPa

psi

Pressure unit is set in the

operating parameters screen.

(see SI unit limitation)

Some data are invalid.

*1:The minimum set of the 1st

press hold setpoint is set to

the initial press setpoint, the

minimum set of the 2nd press

hold setpoint is set to the

1st press hold setpoint, the

minimum set of the 3rd press

hold setpoint is set to the 3rd

press hold setpoint.

1st to 3rd

press hold

time

0

Analysis

cycle

s

Some data are invalid.

SYS Sequential Display Screen

Press F2 (Seq).

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

SYS Sequential Display

GCM #1 / SYS #1 / Method #1 / Stream #1
Time Action
0.0

V1- 1 1st on

5.0

V1- 1 1st off

Main

Gate

Valve

AtmV

DO

Strm#

F1

F2

F3

F4

F5

F6

Figure 5.70

Example of the SYS sequential display screen