<1. Overview>

1-32

IM 11B08A01-01E

Limitation in Selecting a Sampling System

Sample properties that can be conditioned in the GCSMP are as follows. In addition, the
specifi cations for the maximum number of air-operated valves are limited. The sample properties
and specifi cations exceeding these limits necessitate preparing an external sampling system.

Table 1.2

Sample Properties

Sample

Sample valve

Temperature

Pressure

Dust

Mist

Boiling point

Gas

Rotary sampling valve

150°C or less 0.01 to 3 MPa 0.01 g/m

3

(stp) or

less

None

Liquid

Rotary sampling valve

Normal

temperature 0.2 to 3 MPa

None

270°C or less

Liquid sampling valve

with a vaporizer

150°C or less 0.2 to 3 MPa

None

450°C or less

Note: In the case of ATEX, the rotary sampling valve can not be used as the liquid sampling valve.

Refer to Table 1.3 for the limitations on specifi cations with the GCSMP.

Table 1.3

The limitations on specifi cations with the GCSMP

Purpose of air-

operation valve

Number of Mountable Valves

Type 1

Type 2

Automatic stream

switching valve

Max. 7 in total *1

Stream for measurement: 1 to 6

Stream for calibration and

validation: 1 to 3

Max. 4 in total *1

Stream for measurement: 1 to 3

Stream for calibration and

validation: 1 to 3

Atmospheric-pressure

balancing valve

Max. 2

Max. 1

*Max. 1 per 1 GCM

*1

If the number of automatically switched streams exceeds the above limitation, use an external sampling system.

*2

The GCSMP is not mounted on Type 3.

Table 1.4 shows the limitation by the analyzer on specifi cations with an external sampling system

Table 1.4

The limitation by the analyzer on specifi cations with an external sampling system

Automatic stream

switching valve

Type of Stream Valves

Number of

Streams *1

Remarks

One-to-one output (air) *2

1 to 8

One-to-one output (DO) *2

1 to 12

DIO card: 3 ch × 4 (Max. 12 ch)

1 to 20

DO card: 5 ch × 4 (Max. 20 ch)

Binary output (air) *2 *3

1 to 31

4 bits: 1 to 15

5 bits: 16 to 31

Purpose of air-

operation valve

Number of Mountable Valves

Type 1

Type 2

Atmospheric-pressure

balancing valve

Max. 2

Max. 4

*Max. 1 per 1 GCM

*1

Stream for calibration and validation is included.

*2

Air output from the analyzer is 350 kPa.

*3

Only 1 GCM can be set up for an analyzer in the case of binary output (air). (Multiple GCMs are not possible.)

1.11

Human Machine Interface

1.11.1

Types of GC8000 Human Machine Interface

The GC8000 has the following three operation interfaces.

GC-HMI

GC human-machine interface

Displays the status of the GC8000 analyzer, operates it, changes its settings, and

displays analysis data.

12.1-inch color LCD touch panel on the GC8000

ASET

Analyzing server engineering terminal software

PC software which displays status of analyzers and operates analyzers

PCAS

PC analyzer server software

PC software which manages the network and automatically saves data