4 alarm contents, Level 1, Fixed alarm – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 416

Advertising
background image

<7. Troubleshooting>

7-3

IM 11B08A01-01E

7.1.4 Alarm

Contents

The alarm type and contents are as the following tables.
In the alarm display, possible causes and corrective actions are shown. Fix the problems
referring to them.
The thresholds to activate alarms are shown in the table below.
D/I alarm of level 2 and 3 are set from the D/I Setup Screen in EtherLCD.

Level 1

Fixed Alarm

No.

MESSAGE

Alarm Contents

Threshold Item

Threshold

Threshold

Setting

1 SYS ERR

System error

Launch task

Send message

Other

Error to launch task

Error to send

message, etc

N/A

2 OVEN 1 SYS ERR

Oven 1 system error

3 OVEN 2 SYS ERR

Oven 2 system error

4 OVEN 3 SYS ERR

Oven 3 system error

14 DATABASE INIT

ERR

Database init error

Initialize DB

Error to initialize DB N/A

15 CONC CALC ERR

Conc calc error

Calculation time

Calculation was not

concluded before

the start of the next

analysis

Method setting

screen

16 OVEN1 INIT ERR

Oven1 init error

Launch task

Send message

Other – initialize DB

Error to launch task

Error to send

message

Error to initialize DB

N/A

17 OVEN2 INIT ERR

Oven2 init error

18 OVEN3 INIT ERR

Oven3 init error

26 FILE INIT ERR

File initialization error Initialize File system Error to initialize File

system

N/A

27 ETHER INIT ERR

Ethernet initialization

error

Initialize Ethernet

Error to initialize

Ethernet

N/A

28 FTP INIT ERR

FTP initialization

error

Initialize FTP

Error to Initialize FTP

31 CTL CARD ID ERR Card ID error

System setting

Mismatch between

system setting and

card ID

Slot setting

32 OVEN1 CARD ID

ERR

Oven card 1 ID error System setting

Mismatch in

isothermal and

programmed

temperature settings,

and explosionproof

requirements

N/A

33 OVEN2 CARD ID

ERR

Oven card 2 ID error

34 OVEN3 CARD ID

ERR

Oven card 3 ID error

35 TMP CARD1 ID ERR Temp. card 1 ID error System setting

Mismatch in

card marks and

explosionproof types

N/A

36 TMP CARD2 ID ERR Temp. card 2 ID error
37 TMP CARD3 ID ERR Temp. card 3 ID error
38 DET CARD1 ID ERR Detector card 1 ID

error

System setting

Mismatch in card

types

Detector setting

39 DET CARD2 ID ERR Detector card 2 ID

error

40 DET CARD3 ID ERR Detector card 3 ID

error

52 DET1-1 FPGA ERR Detector card 1-1

FPGA error

Write software in

FPGA

Error to write

software in FPGA

N/A

53 DET1-2 FPGA ERR Detector card 1-2

FPGA error

54 DET2-1 FPGA ERR Detector card 2-1

FPGA error

55 DET2-2 FPGA ERR Detector card 2-2

FPGA error

56 DET3-1 FPGA ERR Detector card 3-1

FPGA error

57 DET3-2 FPGA ERR Detector card 3-2

FPGA error

2nd Edition : May 11, 2012-00

7