14 communications setup – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 331

Advertising
background image

<5. EtherLCD>

5-80

IM 11B08A01-01E

5.4.14 Communications

Setup

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Communications Setup

COM # 1
>

- Type DCS

- Protocol MODBUS
- Trans contents * * * * * * * * * *
- Recep contents * * * * * * * *
- Trans speed 9600bps
- Procedure * * * * * * * * * * * * * * * * *
- Parity

Odd

- Com mode RTU mode
- Scaling coef Real

Menu

COM#

F1

F2

F3

F4

F5

F6

Figure 5.88

Examples of communications setup screen

F1 (Menu):

Displays the Table menu screen.

F6 (COM #): Sets a communication port number.

• Modbus protocol can be set to both COM1 and COM2. However, GCCU and GCCU or

Y-Protocol (GC1000/GC8, GC6, GCHC) and Y-Protocol cannot be set to COM1 and COM2
respectively.

• Items that cannot be set due to types and protocols are displayed with * (asterisk).

Type NA

DCS

GCCU

Protocol

×

MODBUS

GC1000/GC8

GC6

GCHC

×

Trans contents

Ч

Ч

Recep contents

Ч

Ч

Trans speed

×

Procedure

Ч

Ч

Parity

×

Com mode

Ч

Ч

Ч

Ч

Ч

Scaling coef

Ч

Ч

Ч

Ч

Ч

For the Y-Protocol, refer to “ Serial Communication (Y-Protocol)” of page App.D-21

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

COM #

1

2

None, DCS, GCCU

Type

○ *

None, DCS, GCCU

Protocol

○ *

MODBUS, GC1000/GC8, GC6,

GCHC * Only when the type is

DCS

Trans

contents

○ *

Analyzed value, Alarm, All data

Recep

contents

○ *

None, Command, All data, GC8

type

Trans speed

○ *

1200 bps, 2400 bps, 4800 bps,

9600 bps, 19200 bps, 38400

bps

Procedure

○ *

Without procedure, With

procedure

Parity

○ *

None, Odd, Even

Com mode

○ *

RTU mode, ASCII mode

Scaling coef

○ *

Real, 9999, 65535

*:

The settings are valid after rebooting or turning off/on its power.

2nd Edition : May 11, 2012-00