Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 228

Advertising
background image

<4. GC-HMI (touch panel)>

4-41

IM 11B08A01-01E

Signal:

Detector signal in mV.

Pressing the

displays the following information:

Input signal:

The analog signal from the detector, which has been sampled every 40
msec, converted to digital values, and then averaged by the given sample
rate

Filtered signal:

Input signal, which has been fi ltered using the fi lter constant set as a
detector signal parameter

Standard deviation of signal:

Standard deviation of input signal for the past 20 inputs

Applied voltage: Bridge voltage of TCD (for TCD only)
Current:

Bridge current of TCD (for TCD only)

Flame detection level:

Set value of fl ame detection level (FID, FID-MC, FPD)

Thermocouple signal:

Thermocouple signal (FID, FID-MC, FPD)

Flame detection status:

Flame detection status (now burning/burning stopped) (FID, FID-

MC, FPD)

Methanizer voltage:

Voltage of methanizer (for FID-MC only)

Temperature information

Indicates the oven temperature information (OVEN, LSV, FPD) of the selected SYS.
Unit: deg C and deg F.

Temperature controller information

Indicates the SV value, PV value, and ON/OFF status for the isothermal oven, LSV, and

FPD, of the selected SYS.

Press

.

Figure 4.54

Pressure information

Displays pressure information of EPC (Carrier n-1 to n-2, and Utility n-1 to n-4 (where n is

oven number)) inthe selected oven.

2nd Edition : May 11, 2012-00