Calibration set – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 325

Advertising
background image

<5. EtherLCD>

5-74

IM 11B08A01-01E

Display Item

Name

Set Lower Limit Upper Limit

Unit

Remarks

GCM #
Cal (Val)

method

Manual, Semi-auto, Auto

Auto start

date

2000 to 2099, Jan to Dec, 1st

to 31st*

* Upper Limit will differ

depending on the month

Auto start

time

0 to 23 hours, 0 to 59 minutes

Time interval

0 to 255 days, 0 to 23 hours, 0

to 59 minutes

Date, Hour, and Min cannot be

set to all 0.

Auto cal

status

Stopping, Executing

Calibration set

Press F4 (Cal).

1 1 / 1 1 / 2 2
1 5 : 1 5 : 4 5

Calibration Setup

GCM #

1 Cal # 1

>

- Cal stream 2

- Cal times 1
- Auto cal Unexecuted
- ValStr# before Cal 0
- ValStr# after Cal 0
- Val before Cal times 1
- Val after Cal times 1

Menu

Peak

Main

GCM#

Cal#

F1

F2

F3

F4

F5

F6

Figure 5.82

Example of calibration set screen

F1 (Menu):

Displays the Table Menu screen.

F2 (Peak):

Displays the calibration peak set screen.

F4(Main):

Displays the Cal/Val Set (main) screen.

F5 (GCM #): Specifi es a GCM number.
F6 (Cal #):

Specifi es a Cal number.

When ValStr # before Cal and ValStr # after Cal are set to 0, Val stream is not executed.

• When a GCM number is changed, the Cal number is allowed to return to 1.
• A Cal stream can be set to Cal only when a stream type is set to Cal. When the setting is

attempted with a stream type other than Cal, the message “The kind of stream is incorrect”
is displayed. When it does not exist in the relevant GCM, the message “The stream doesn’t
exist in GCM” is displayed.

• A ValStr # before Cal and a ValStr # after Cal can only be set when the stream type is Val.

When the stream type is other than Val, the message “The kind of stream is incorrect” is
displayed. When it does not exist in the relevant GCM, the message “The stream doesn’t
exist in GCM” is displayed.