<6. Maintenance>

6-51

IM 11B08A01-01E

L9805ZF

Regulator valve for samples (Diaphragm: Tefl on, Liner: SUS)

L9862AB Filter

Needle valve

Stream switching valve

Regulator for samples

Filter

Cap nut

Nut

Figure 6.44

Example of Sample Processing System (May contain a Stop Valve depending on the

system)

Filter element

(1) Stop the operation. (See “3.3.4 Stopping operation” for this procedure).
(2) Turn off the supply of the sample.
(3) If the sample is liquid, purge the sample line with the purge gas (nitrogen gas or instrument

air).

• When the stream switching valve is a pneumatic valve, change the analyzer status to

Manual and turn on the stream switching valve on the sample line to let the purge gas fl ow
in.

• When the stream switching valve is a stop valve, close all the stop valves and open the stop

valve on the sample line to let the purge gas fl ow in.

(4) Close all the stream switching valves.

IMPORTANT

• During the purge, purge gas may fl ow into the sample return line, causing pressure

fl uctuation.

Prepare the sample return line and collection tank, if necessary.
• Prepare the sample return line so that the sample does not return.

CAUTION

The isothermal oven is extremely hot after turning off the power immediately. Keep the purging
air supplied for more than an hour after turning off the power. Keep hands away from the oven
components.

(5) Remove the cap nut (see Figure 6.45).
(6) Replace the internal fi lter element and the gaskets with a new one.
(7) Firmly tighten the cap nut.
(8) Confi rm that there are no leaks.

<Replacement Parts>

L9862AG Filter

element

L9862AC Gasket