Caution – HT instruments MACROTEST G3 User Manual

Page 19

Advertising
background image

MACROTESTG3 - COMBIG3

EN - 18

CAUTION

 Disconnect any cable not strictly involved in measurement and especially

check that no cable is connected to In1 input.

 Before connecting the test leads, make sure that there is no voltage at the

ends of the conductors to be tested.

6. Connect the alligator clips and/or test leads and/or remote lead to the ends of the

conductors to be tested as in Fig. 8 and Fig. 9.

7. Press the GO/STOP key on the instrument or the START key on the remote lead. The

instrument will start the measurement.

CAUTION

During this whole stage, do not disconnect the test leads of the instrument
from the conductor under test. It could remain charged with a dangerous
voltage due to the stray capacitances in the circuit being tested.

8. Regardless of the operating mode selected, at the end of the measurement, the

instrument applies a resistance to the output leads to discharge the capacitances in the
circuit.

9. In mode:

The final result is the minimum insulation value measured during the test

Pressing a second time the GO/STOP key or the START key on the remote lead

stops the test before the set time has elapsed.

10. The measurement result is shown both as a numeric

value and in the analog bar graph as shown in the
screen to the side. The values of the real test voltage
and the measurement time are present on the display.

The

symbol indicates the positive result of the

measurement.

Press the SAVE button or touch the

icon to save

the measurement (see § 7.1).

11. At the end of the test, if the value of the measured

resistance is lower than the set limit, the screen to the
side is shown on the display.

The value is shown in red and the

symbol indicates

the negative result of the measurement.

Press the SAVE button or touch the

icon to save

the measurement (see § 7.1).