HT instruments MACROTEST G3 User Manual

Page 56

Advertising
background image

MACROTESTG3 - COMBIG3

EN - 55

8. At the end of the test, if the detected phase sequence is

correct, the instrument displays a screen like the one
shown to the side (result "1-2-3").

Press the SAVE button or touch the

icon to save

the measurement (see § 7.1).

9. At the end of the test, if the two detected voltages are

in phase (phase concordance between two distinct
three-phase systems
), the instrument displays a
screen like the one to the side (result "1-1-").

10 At the end of the test, if the detected phase sequence is

not correct, the instrument displays a screen like the
one shown to the side (result "2-1-3").

Press the SAVE button or touch the

icon to save

the measurement (see § 7.1).


Advertising