HT instruments MACROTEST G3 User Manual

Page 29

Advertising
background image

MACROTESTG3 - COMBIG3

EN - 28

6.3.6. Anomalous situations
1. If the voltage between inputs B1 and B4 and inputs B1

and B3 is higher than 265V, the instrument provides the
warning screen shown to the side and blocks the
execution of the tests.

2. If the voltage between inputs B1 and B4 and inputs B1

and B3 is lower than 100V, the instrument provides the
warning screen shown to the side and blocks the
execution of the tests.

3. If the instrument detects the absence of the signal to

terminal B1 (phase conductor), it provides the warning
screen shown to the side and blocks the execution of
the tests.

4. If the instrument detects the absence of the signal to

terminal B4 (neutral conductor), it provides the warning
screen shown to the side and blocks the execution of
the tests.

5. If the instrument detects the absence of the signal to

terminal B3 (PE conductor), it provides the warning
screen shown to the side and blocks the execution of
the tests.