7.4

PerForMinG MeASUreMentS – all users

Place the zero solution in the sample chamber and close the lid.

Selecting the

Zero key will start the cycle, setting zero at each wavelength in turn. A series of

information boxes will advise of the progress.

When completed all displayed absorbance values will be set to zero (or 100% Transmittance).

Replace the zero solution with the sample to be measured and close the sample chamber lid.

Selecting the

Read key will start the measure cycle, measuring at each wavelength in turn. A series of

information boxes will advise of the progress.

When completed the measured absorbances will be displayed.

The results of the selected calculation will be displayed below the Wavelength/Absorbance data.

Saving a method allows access to the table through

Recalling a Method.