Jenway 6715 User Manual

Page 165

Advertising
background image

147

Pump tube Schematic

traditional Pumping Arrangement

Checking the Sipper Pump

After fitting the tubing it is recommended that the following check is performed to ensure the cuvette
fills correctly and that there are no leaks present.

Select

Photometrics from the Main Menu. (You do not need to log in for this check).

Select

Settings and the icon.

Select

Independent from the Operating Mode options.

Press to accept.

The switch on the front of the accessory will illuminate.

Offer up a receptacle of water or coloured solution to the inlet tube. Press the accessory switch
and allow the pump to run, ensuring the cuvette fills correctly and that no leaks are present. If the
operation is satisfactory press the accessory switch again to stop the pump.