important: When fitting the replacement lamp it is essential that the glass is not touched. Finger
marks will damage the lamp. Should accidental damage with finger marks occur, the surface of the
lamp can be cleaned using iso-propyl alcohol.

Gently insert the lamp module into the housing ensuring it is fully pushed home.

Refit the 4 Flash Lamp Retaining Screws. Plug in the Flash Lamp Connector.

Insert the complete assembly into the base plate and refit the 2 posi-head Lamp Module Retaining
Screws.

Carefully turn the instrument over. Reconnect the mains supply to the instrument and power up. The
self-test power up sequence will be initiated. On successful completion the instrument is ready for use.

In the unlikely event of further failure on power up self-test, please consult the manufacturer or you
local dealer for advice.