Jenway 6715 User Manual

Page 107

Advertising
background image

89

Setting Start level

Selecting the

Start Level key displays a numeric keypad. Clear the current setting and enter the

required level above or below which data collection is required.

Limits for this setting are – 0.299 to 2.999Abs and 0.1 to 199.1%T.

Setting Method Security - This option is only applicable to logged in users.

Setting Concentration Parameters

The rate of change of absorbance can be converted to concentration using a known factor or
reference concentration value. Whichever is not required should be set to unity so that it is ignored in
the calculation.

Standard

Select the Standard key and enter the concentration value of the calibration standard. Values from

–9999.99 to 9999.99 can be entered. If a concentration value is not to be used then this should be set
to 1.00.