Jenway 6715 User Manual

Page 106

Advertising
background image

88

Setting run time

Selecting the

Run Time key displays a numeric keypad. Clear the current setting and enter the

required time, in seconds, for the measurement period. The limits for this setting are from 30 to 9999
seconds (2.75 hours).

Setting lag time

Selecting the

Lag Time key displays a numeric keypad. Clear the current setting and enter the

required lag time in seconds. The lag time is a portion of the run time over which no data is collected,
starting from the beginning of the run time. This allows for any latency or non-linear reaction to be
disregarded. Setting this to zero means data will be collected over the full run time. The limits for this
setting are from 0 to 9999 seconds.

Setting Start on level

Selecting the

Start On Level key disables the lag time option. Repeated presses of this key cycle

through the

Greater Than, Less Than and Disabled options. With Greater Than or Less Than

selected data collection will start when the measured value (Absorbance or Transmittance) is above or
below the value entered in the

Start Level setting.