Jenway 6715 User Manual

Page 28

Advertising
background image

10

1.2

inStrUMent SPeCiFiCAtionS

technical Specification

6700

6705

6715

Light Source

Tungsten Halogen Xenon

Xenon

Spectral Bandwidth

4nm

4nm

1.5nm

Stray Light

<0.1% at 340nm <0.05% at 220nm <0.05% at 220nm

Wavelength Range

320-1100nm

190-1100nm

190-1100nm

Common Specifications
Optics

Sealed, MgF Coated, Split Beam

Wavelength Resolution

0.1nm

Wavelength Accuracy

±1.0nm

Wavelength Repeatability

±0.2nm

Photometric Ranges

-0.300 to 3.000A & 0 to 199.9%T

Photometric Resolution

0.001A & 0.1%T

Photometric Accuracy

±0.005 at 1A

Photometric Stability

<0.001A per hour

Quantitation Range

-99999 to +99999

Number of Standards

20 with up to 5 replicates of each

Curve Fit Algorithms

Linear, Quadratic and Cubic Functions

Multi-wavelength Data Points

Up to 4 wavelengths

Calculations

Difference and ratio

Kinetics Time Limits

30 to 9999 seconds

Kinetics Calibration

Standard or factor

Scan Speed

1500nm/minute at 0.1nm data steps

Post Scan Analysis

Peak/Valley pick, Peak Ratios, Area, Zoom, Wavelength Table,

Derivatives, Smoothing

Configuration

Secure Multi-User and Free Access

Number of Users

10 + Supervisor

Number of Methods

>500 (on internal flash memory or removable media)

Results Storage

>500 (on internal flash memory or removable media)

Removable Media

MM/SD Memory Card or SD/USB memory card

Interface

USB, Centronics, Analogue

PC Software

Supplied on CD-ROM with USB interface cable

Mains Supply

100 to 230V ac 50 or 60Hz

Sample Compartment

150mm (max. height) x 130mm (w) x 210mm (d)

Size

490x390x220mm

Weight

7.5Kg

Advertising