Jenway 6715 User Manual

Page 81

Advertising
background image

63

6.4

PerForMinG MeASUreMentS – all users

Performing a Scan

Once the required settings have been confirmed the main

Spectrum measurement screen will be

shown with the

Scan key inactive (blanked out):

Performing a Baseline

Insert a cuvette containing a blank solution into the sample holder and close the sample chamber lid.
(Test tubes or other sample containers may be used depending on the sample holder accessory fitted).
Select the

Baseline key. The messages ‘Performing baseline scan. Please wait… are displayed

momentarily. Once the baseline measurement and storage is completed the screen will update to
show:

note: In general the blank solution should contain everything that is in the sample except the
colour-producing component. For specific information reference should be made to the procedure or
application being followed. For enhanced reproducibility matched cuvettes should be used.

Scanning the Sample

Once the baseline has been performed the

Scan key will be active. Remove the cuvette containing the

blank solution and insert the sample into the sample holder and close the sample chamber lid. Select
the

Scan key, if Auto Scale has been selected once the scan is completed, the screen will blank and

the scan will be re-drawn to the optimum scaling.