154

Flash lamp Module replacement – Models 6705 and 6715

Warning: Disconnect the instrument from the mains supply prior to replacing the lamp. Care should
be taken when removing the module. Ensure the module is cool prior to handling.

Access to the lamp module is gained via the base of the instrument.

Safety glasses must be worn when UV emissions are present.

important: Prior to turning the instrument upside down the sample chamber must be
checked and any samples removed to avoid spillage or damage.

Carefully turn the instrument over onto a protected work surface and locate the

2 posi-head Lamp

Module Retaining Screws and place carefully to one side, (refer second diagram).

no attempt should be made to remove the 2 socket head screws from the base.

Gently withdraw the housing from the base plate and unplug the Flash Lamp Connector.

Remove the 4 Flash Lamp Retaining Screws and place carefully to one side. Locate and carefully
remove the flash lamp module.

Carefully remove the replacement lamp module from the packaging, ensuring the glass portion of the
lamp is not touched.

Lamp module

retaining screws

Flash lamp

retaining screws

Flash lamp

connector