Change the displayed harmonics orders, Graph”/ ”list, Rate of content”/”phase angle”/ rms value (power) – KYORITSU 6315 Instruction Manual User Manual

Page 115: V”/”a”/”p/σp

Advertising
background image

Displaying the list of harmonics

KEW6315

- 113 -

KEW6315

”Change the displayed harmonics orders”

Press the

Key to scroll the page vertically.

”Graph”/ ”List”

Press the

Key to display voltage/ current/ power harmonics, from 1

st

to 50

th

order, in list or

graphic form. Only the rate of harmonics content can be checked on graph display screen,

”Rate of content”/”Phase angle”/ RMS value (Power)”

Press the

(Rate/ DEG/ RMS) Key to change the displayed items on the list. While “V”:voltage or

“A”: Current are displayed on the screen, Rate/ DEG (phase angle with V1 basis (0

o

)) / RMS are switchable.

While “P” (ΣP): Power is displayed, Rate/ DEG (voltage/ current phase angle per ch) / Power are

switchable.

“V”/”A”/”P/ΣP”

Press the

(V/A/P/ΣP) Key and select the items to be analyzed: V: voltage/ A: Current/

P: Power (ΣP: Sum per system, Total amount).