11.2 Error messages and actions

KEW6315

- 159 -

KEW6315

the instrument or the sensor.

Message

Detail & Action

Battery level is low.

Powering off

● Connect the instrument to an AC power source, or replace the

batteries with new ones. * Size AA Alkaline battery (LR6) or

fully-charged Size AA Ni-MH battery x 6pcs

See “How to install batteries” (P. 31).

Not having free space on the internal

memory. Format the memory or

delete unnecessary files.

● Check the free space on the internal memory and the number of

the saved files. Max number of the file that can be saved on the

memory is: 3 for measurement data and 8 for the other data. If

the free space is not enough, delete unnecessary files, format

the memory.

See “To delete, transfer or format the recorded data

(P.

82).

Cannot read the setting file. The file

may be damaged.

● Try again. If still the setting files are not read;

* problems with SD card or KEW6315 are suspected, if the setting

files are on the SD card,

* problems with KEW6315 are suspected, if the setting files

are in the internal memory.

If the problem with KEW6315 is suspected, stop using the

instrument.

Available memory is low.

Check the amount of free space in

the SD card and internal memory.

● Check the free space and the number of saved files on the SD

card and the internal memory. Max number of the file that can be

saved on the memory is: 3 for measurement data and 8 for the

other data. If the space is not enough, delete unnecessary files,

format the card or memory. When using another SD card, it

should be formatted on KEW6315, not on the PC.

See “To delete,

transfer or format the recorded data” (

P.

82).

There is no available space in the

storage area.

Start time is set in the past.

Check the recording start method.

● REC Start is either “Constant rec. / Time period rec.” and the time

set for REC End is set to the past. Check and modify the time and

date. See “(8)/ (9) Setting for recording method” (P. 45).

Failed to start recording.

● Check the “Recording setting” at SET UP menu.

See “

5.4 Recording setting” (P. 71).

● Try again. If still a record does not start, there may be a problem

with either the SD card or the internal memory. Check which is set

as the destination to save the data. If the destination is internal

memory, a problem with KEW6315 is suspected. Stop using the

instrument in this case.

Cannot change the instrument

settings during recording or in

stand-by mode.

● Change of setting is not allowed during a record. To change the

settings, stop record and confirm “Recording stopped.” message

appears and then disappears.

Advertising