Connection to input/ output terminals

KEW6315

127

KEW6315

1 Open the Connector cover.

2 Press the rectangular protrusion above a terminal with a flat-blade screw driver, and insert

a signal wire.

3 Remove the driver and fix the wire.

”Input terminal”

For monitoring the voltage output signals of Thermo sensors. These terminals are useful to measure the

signals from other devices and power failures at the same time.

Number of Ch: 2ch

Input resistance : approx 225.6kΩ

”Output terminal”

For fixing the generating outputs to “Low” while power quality events are lasting. Usually, it is fixed to “High”,

but changed to “Low” if the duration of an event is less than 1 sec. This is applicable to the events with the

highest-priority only. To adjust the generating outputs to the events with low-priority, select “OFF” for the

events with higher priority than the desired event. The details are described in “Threshold setting for Power

quality (Event)” (P. 65). * [Priority order]: Transient -> INT -> Dip -> Swell -> Inrush current

Output format : Open collector output

Max input : 30V, 50mA, 200mW

Output voltage : Hi – 4 to 5V

Lo – 0 to 1V

Signal wire

+5V

H

L

OUTPUT

Flat-blade screwdriver