Items measured at Harmonics measurement

KEW6315

- 144 -

Items measured at Harmonics measurement

Meas. system

: Digital PLL synchronization

Meas. method

: Analyze harmonics, and then add and display the inter-harmonics

components adjacent to the integral order of the analyzed harmonics

Effective frequency range

: 40 - 70Hz

Order analysis

: 1 - 50th

Window width

: 10-cycle at 50Hz, 12-cycle at 60Hz

Window type

: Rectangular

Data analysis

: 2048 points

Analyzing rate

: once/ 200ms at 50Hz/60Hz

R. m.s. harmonics voltage Vk [Vrms]

Range

Same as r.m.s. voltage

Displayed digit

Same as r.m.s. voltage

Display range

Same as r.m.s. voltage

* rate of content 0.0% - 100.0%, percentage against the basic wave

Measurement

system

Complied with IEC61000-4-30, IEC61000-4-7, IEC61000-2-4

Analysis window width is 10/12-cycle for 50/60Hz, and the measured values

contain the inter-harmonics components adjacent to the analyzed order.

Accuracy

Complied with IEC61000-2-4 Class3 where 10% - 100% of input range for

600V Range.

3% or more against 100V of nominal voltage

: ±10%rdg

Less than 3% against 100V of nominal voltage : nominal voltage ±0.3%

1000V Range

: ±0.2%rdg±0.2%f.s.

Equation

(

)

(

)

(

)

(

)

=

+

+

+

=

1

1

2

2

10

10

n

i

n

k

c

r

n

k

c

ck

V

V

V

1

100

c

ck

V

V

×

=

c: Measurement channel, k: Harmonics of each order
Vr: Real number after Voltage FFT conversion
Vi: Imaginary number after Voltage FFT conversion
Measurement cycle in this equation is 10-cycle. For 12-cycle measurement,

"10k+n" should be replaced with "12k+n”.