Ng judgment wiring check, Causes – KYORITSU 6315 Instruction Manual User Manual

Page 45

Advertising
background image

Start measurement with “Quick start guide”

-

NG judgment

Wiring check

Close the result display. Then, the blinking vectors and the

values of NG items will be displayed. If

ideal vector diagram will be displayed

Criteria of judgment and cause

Check

Criteria of Judgment

Frequency

Frequency of V1 is within 40
- 70Hz.

- Voltage clip is firmly connected to the DUT?
- Measuring too high

AC Voltage

input

AC voltage input is 10% or
more of (Nominal voltage x
VT).

- Voltage clip is firmly connected to the DUT?
- Voltage test lead
AC v

Voltage

balance

AC voltage input is within
±20% of reference voltage
(V1).
* (not checked for single-

phase wiring)

- Settings are matched with the wiring system under
test?
- Voltage clip is firmly connected to the DUT?
- Voltage test lead
AC v

Voltage

phase

Phase of AC voltage input is

within ±10º of reference value

(proper vector).

- Voltage test leads are properly connected?

(Connected to proper channels?)

Current

input

Current input is 5% or more

and 110% or less of (Current

Range x CT).

- Clamp sensors are firmly connected to the Power

input terminals on the instrument?

- Setting for Current Range is appropriate for input levels?

Current

phase

- Power factor (PF, absolute

value) at each CH is 0.5 or

more.

- Active power (P) at each

CH is positive value.

- Arrow mark on the Clamp sensor and the orientation

of flowing current coincide with each other?

(Power supply to Load)

- Clamp sensors are connected properly?

KEW6315

- 43 -

KEW6315

Close the result display. Then, the blinking vectors and the

values of NG items will be displayed. If everything is OK, the

ideal vector diagram will be displayed in the lower left corner.

Causes

Voltage clip is firmly connected to the DUT?
Measuring too high harmonic components?

Voltage clip is firmly connected to the DUT?
Voltage test lead is firmly connected to the

AC voltage input terminal on the instrument?

Settings are matched with the wiring system under

test?

Voltage clip is firmly connected to the DUT?
Voltage test lead is firmly connected to the

AC voltage input terminal on the instrument?

Voltage test leads are properly connected?

(Connected to proper channels?)

Clamp sensors are firmly connected to the Power

input terminals on the instrument?

Setting for Current Range is appropriate for input levels?

Arrow mark on the Clamp sensor and the orientation

of flowing current coincide with each other?

(Power supply to Load)

Clamp sensors are connected properly?