KEW6315

- 123 -

KEW6315

Chap. 7 Other functions

“Data hold”

Display update can be disabled by pressing the “DATA HOLD” Key. The “

” icon will appear while

display update is disabled. The icon will disappear and display update will be enabled by pressing the

“DATA HOLD” Key again. Switching screens is possible, moreover, measured values and event information

are continuously recorded even while the Data hold function is activated.

“Key lock”

Pressing the “DATA HOLD” Key 2 sec or more disables all Keys, except for LCD key, and “ ” icon

appears. Another long press (2 sec or more) is required to restore the disabled Keys.

“Turning off the Backlight”

Press the LCD Key to turn off the backlight. Pressing any keys, except for the Power key, turns on the

backlight again.

“Backlight Auto-off”

While KEW6315 is connected to an AC power source:

The LCD backlight is turned off automatically 5 min after the last key operation. Press any key except

for the Power key to turn on the light again. To disable the Backlight auto-off function, select “Disable

auto-off” on the setup menu.

While KEW6315 operates with battery:

The brightness will be cut by half. The backlight will be automatically turned off 2min after it is turned on.

Press any key except for the Power Key to turn on the backlight again. The backlight does not on

continuously while the instrument is operating with batteries.

“Auto-power-off”

While KEW6315 is connected to an AC power source:

The instrument is powered off automatically 5 min after the last key operation. This function does not

operate while the instrument is recording data. Press the Power key to power on the instrument again.

To disable the auto-power-off function, select “Disable auto-off” on the setup menu.

While KEW6315 operates with battery:

The instrument is powered off automatically 5 min after the last key operation. This function does not

operate while the instrument is recording data. Press the Power key to power on the instrument again.