Section 3 - Electrical Connections and Setup

Page 8

S

H

A

R

K

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

Section 3 - Electrical Connections and Setup

3.4 pH or ORP Combination Probe connections and setup

The drawing shows the connections for the Aquametrix Combination probe. The cable should be run in
a conduit separate from AC power wires, and via a separate conduit hole. The cable length should not
exceed 10 feet (3 meters).

These wires are only
present with 4-wire
combination probe.

JUMPER 2-4 MUST BE
INSTALLED FOR
COMBINATION PROBE
(CUSTOMER SUPPLIED)

COAX SHIELD (4)

COAX CENTER (3)

WHITE (5)

GREEN (4)

P6

8

1 2 3 4

6

5

7

10

9

11 12 13 14 15 17

16

COMBINATION
pH or ORP PROBE

Fig. 3.3 Connections for the 2 and 4 wire Combination Probe

SEC. 5.8

SEC. 5.5

SEC. 5.1

ORP PROBE

SELECT COMBINATION PROBE

SELECT ORP METER

MANUAL CALIBRATE

PH PROBE

MANUAL CALIBRATE

SELECT PH METER

SELECT COMBINATION PROBE

SEC. 4.1

SEC. 4.6

SEC. 4.7

SEC 4.10

MANUAL PROBE
TEMPERATURE SETUP

DISABLE TEMPERATURE

SEC 5.10

DISPLAY

MANUAL CAL PH

METER SELECTION

PROBE SELECT

7.15pH 25.0C

T.COMP OVERRIDE

7.15pH 25.0C

PROBE SELECT

METER SELECTION

MANUAL CAL PH

T.DISP OVERRIDE

RUN MODE

TWO WIRE PROBE WITH NO TEMP SENSOR

TWO WIRE PROBE WITH NO TEMP SENSOR

RUN MODE

ORP

pH

Once connected, step throught the LCD menus to select the probe in the order
shown. The first two steps may be skipped if the meter is already configured for
a Combination Probe. If a two wire pH probe is used, which has no temperature
sensor, ensure that the Temp. Comp. Override is set to same temperature as
the buffer before calibrating. If a two wire ORP probe is used, you can blank the
Temp display with the T DISP OVERRIDE menu.

The

2 wire version has no

temperature sensor and is
connected via a coaxial wire.
In a pH meter, the user
should set the

T COMP

OVERRIDE

menu to

ON

(Section 4.11) and adjust the
temperature setting to the
actual probe temperature.
In an ORP meter, the user
should set the

T.DISP OVER-

RIDE to ON (Section 5.10) to
blank the temperature reading
on the display.

The

4 wire version has two

additional wires for the probe
internal temperature sensor.
Ensure that the

T COMP

OVERRIDE or T.DISP
OVERRIDE
is OFF.

Note: Leave 4” to 6” slack for
all wires connected to the ter-
minals of P6. Slack required
so wires do not interfere with
opening/closing of front door.

Dwg# N104-35

CAUTION:
Always remove line
power before unplugging
or plugging in the P6
connector