3.7 Relay connections

The Shark controller has three internal relays. Relays A and B are for control, the Alarm Relay
can be configured for alarm functions or as an additional control relay.

The connections to the relays are shown in the drawing.
Note that the AC power is internally connected to the
relay terminal plug P4. This is used to provide 120V or
240V AC power for the relay contacts.

Wire Specification: Size and fuse wire accroding to local
electrical code. Wire size not to exceed 14 AWG.

S

H

A

R

K

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

Section 3 - Electrical Connections and Setup

Section 3 - Electrical Connections and Setup

Page 11

WARNING

DISCONNECT POWER FROM

CONTROLLER AND LOADS WHILE

CONNECTING TO THE RELAY

OUTPUT TERMINAL PLUG.

N.C.

N.O.

N.C.

N.O.

N.C.

N.0

GND

RELAY B

ALARM

RELAY A

NEUTRAL

GROUND

LINE

LINE OUT (21)

connections

Internal

NEUTRAL OUT (22)

RELAY CONNECTIONS FOR 120VAC

SHARK-120

P3

G

L

N

21

P4

22

23

24

25

26

27

28

29

30

31

W A R N I N G

PIN 21 must be
connected to relays
for power.

PIN 22 cannot be
connected to the
relays.

N.C.

N.O.

N.C.

N.O.

N.C.

N.0

GND

RELAY B

ALARM

RELAY A

RELAY CONNECTIONS FOR 240vac

SHARK-240

LINE

GROUND

LINE

LINE OUT (21)

connections

Internal

LINE OUT (22)

G

L

L

P3

21

P4

22

23

24

25

26

27

28

29

30

31

W A R N I N G

PIN 21 must be
connected to relays
for power.

PIN 22 cannot be
connected to the
relays.

Figure 3.6 Connections for Relay A, B and Alarm

Caution:

The contacts are rated at 10 amp N.O. and 5 amp N.C. Do
not exceed this rating. When switching larger currents, use an
auxillary relay switched by the controller relay to extend the
controller relay life. If the relays are controlling an inductive
load, use appropriate transient suppression at the load.

Dwg# N104-38