MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

pH - Utilities Menu - Manual Test Mode 4.4

pH - Utilities Menu - Manual Test Mode 4.4

Page 22

DOWN

DOWN

DOWN

UP

DOWN

DOWN

UP

DOWN

Manual Test Mode is used to simulate a process reading
in order to verify the correct response of the outputs.
When in the Manual Test Mode, the outputs are no
longer placed on hold as they are when in the rest of the
menu.

NOTE:

When the user exits the Manual Test Mode, the relays

and 4-20mA outputs will remain in the Test Mode state

until the operator enters the run mode. The relays and

outputs will then revert back to the previous On-line state.

The CH1 4-20 mA output will
track the change in process

When a relay is energized, the
characters a,b,c will change to
upper case A,B,C

RUN MODE

UP

NOTE

PRESS THE AND KEYS

TOGETHER TO GO IMMEDIATELY BACK TO

RUN MODE

DOWN

As the value is changed, the state of the relays
will change depending on their settings in the
OUTPUTS menu.
a/A - state of Relay A
b/B - state of Relay B
c/C - state of Alarm relay

Press the key once to
move the cursor to the RH side
of the display

As well, Channel 1 4-20 ma output will also
follow the process value change. The actual
change will depend on how the CH1 output
was scaled in the OUTPUTS menus.
(See Section 4.21)

Press the key to return to the
MANUAL TEST MODE menu

Use the and keys to change
the simulated process value.

Press the key once which will
move the cursor over the least digit of the
simulated process value.

TEST 7.0 0 >

TEST 7.00 >

7.15pH 25.0C

CALIBRATION

UTILITIES

a b c 19.0mA

a b c 19.0mA

TEST 13.1 5 >

TEST 13.15 >

MANUAL TEST MODE

a b c

12.0mA

a b c

12.0mA

a B c

19.0mA

TEST 13.1 5 >

DOWN