Orp - led display menu - orp manual calibrate 5.22, Manual calibrate – Metex Shark Multi-Parameter Controllers & Analyzers User Manual

Page 76

Advertising
background image

ORP - LED Display Menu - ORP Manual Calibrate 5.22

Page 74

S

H

A

R

K

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

ORP - LED Display Menu - ORP Manual Calibrate 5.22

Press the key to go
immediately back to Run mode

CAL

CAL

DOWN

ENTER

ENTER

ENTER

ENTER

UP

DOWN

UP

ENTER

ENTER

CAL

RUN MODE

MANUAL CALIBRATE

ENTER

ENTER

The Controller will display bUFF, prompting
the user to place the probe in the buffer
solution.

Be sure to clean and rinse the Probe first with
D.I. water and then insert it in the Buffer.
Press the key.

Use the and keys to adjust
the display until it reads the correct value of
the buffer.

Press the key to accept the setting.

Press the key again to confirm the
change in calibration value.

The controller will flash "bUFF" for about 10
seconds while is reads the buffer. It will then
display the mV value.

The display will flash

MANUAL CALIBRATE
This menu allows the
user to manually
calibrate the probe, the
user enters the known
mV buffer value

Press the
key to select
Manual Calibration