Conductivity - Calibration Menu - Dry Cal Cond 6.2

Conductivity - Calibration Menu - Dry Cal Cond 6.2

Page 78

Dry Calibration eliminates the need for conductivity reference
solutions, the user inputs the Cell K factor supplied by the
factory.

If the conductivity cell has a tag attached to it, specifying the
exact cell constant, the user is prompted to enter this value.

DOWN

DOWN

DOWN

UP

DOWN

DOWN

UP

NOTE

PRESS THE AND KEYS

TOGETHER TO GO IMMEDIATELY BACK TO

RUN MODE

DOWN

Press to store the
calibration data and return
back to the Dry Cal Cond
menu so the user can select
another function.

If the calibration did not appear
to be correct, press the
key which will return back to the
Manual Cal Cond menu.

DOWN

NOTE:
ACCELERATOR KEYS
Pressing the or
key once will change the value
by the smallest digit. Holding
the key down will cause the
value to change at an
increasing rate until the key is
released. Pressing the key
again will cause the value to
start changing at it's slowest
rate again. This allows the user
to get to the new multiplier
value quickly.

DOWN

UP

If the setting is OK, use the
key to move the cursor over the Y
text and press the down key.

DOWN

UP

Then press to move
the cursor to the RH position

Use the and keys to
adjust the value to the K factor
specified on the probe tag

Press the key once
which will move the cursor
over the least sign. digit of
the display.

DOWN

RUN MODE

CALIBRATION

K FACTOR 10.000 >

K FACTOR 10.00 0 >

DRY CALIBRATION

DRY CALIBRATION

DRY CALIBRATION

STORE? Y N

STORE? Y N

K FACTOR 1.000 0 >

DRY CALIBRATION

K FACTOR 1.0000 >

DRY CALIBRATION

MANUAL CAL COND

DRY CAL COND

DRY CALIBRATION

1000uS 25.0C